FROST (Fridge Remote Operations, Status, and Thermometry)

Primary cryostat control software for MEC' (MKID Exoplanet Camera) Prime

Overview

FROST provides a single interface for controlling and monitoring all cryostat hardware via serial communication. It exposes both a command-line interface for scripting and direct device control, and a desktop GUI for interactive operation.

Device Role
Lakeshore 625 Superconducting Magnet Power Supply — ramps/de-ramps ADR magnet
Lakeshore 370 AC Resistance Bridge — monitors device stage temperature, controls PID loop with LS625
Lakeshore 350 Temperature Controller — all stage and GL7 thermometers, GL7 heater and switch outputs
Heatswitch Zaber T-NM17A04 stepper motor — opens/closes heat switch during ADR ramp
Compressor Cryomech compressor — turns PT cooling on/off during cooldown

Requirements

  • Rust 1.70+
  • GUI environment (X11 or Wayland) for the GUI mode
  • cryomech_api crate at ../cryomech_api/ (path dependency — see Dependencies)

Installation

After building, the binary lives at target/release/frost. The recommended installation is a symlink because it survives rebuilds without any extra steps:

sudo ln -sf $(pwd)/target/release/frost /usr/local/bin/frost

Only needs to be run once. Every subsequent cargo build --release automatically updates the installed binary. This enables the user to use FROST functionality from any directory without being inside the project directory.

Verify the installation:

frost --help

Building

# Build optimized release binary
cargo build --release

# Run tests (no hardware required — all tests use nonexistent port paths)
cargo test

# Run hardware-dependent tests (requires physical devices connected)
cargo test -- --include-ignored

Serial Port Configuration

All device ports have hardcoded defaults that match the MEC Prime set-up. They can be overridden per-command with --port and --baud flags.

Device Default Port Default Baud
Lakeshore 625 /dev/ttyUSB0 9600
Lakeshore 370 /dev/ttyUSB1 9600
Lakeshore 350 /dev/ttyUSB2 57600
Compressor /dev/ttyUSB3 115200
Heatswitch /dev/ttyUSB4 9600

To use a different port for a single command:

frost lakeshore625 --port /dev/ttyUSB5 current
frost lakeshore370 --port /dev/ttyUSB2 --baud 1200 kelvin 1
frost heatswitch --port /dev/ttyUSB1 --device-id 2 open
frost compressor --port /dev/ttyUSB0 --addr 17 status

There are no config files. If you need permanent port reassignment, update the DEFAULT_PORT constants in the relevant src/<device>.rs file.

Busy-port retry behavior

If a serial port is temporarily held by another process, FROST retries the open before failing:

  • SCPI ports (Lakeshore devices): up to 10 retries, 15 seconds apart (~2.5 minutes max)
  • Zaber port (heatswitch): up to 2 retries, 15 seconds apart (~30 seconds max)

If retries are exhausted, the command fails with an error naming the port.


GUI

Launch the interactive GUI:

frost gui
# or, to build and run in one step:
cargo run --release -- gui

The GUI runs a background worker thread that polls all devices every 30 seconds (staggered to avoid serial bus contention) while keeping the interface responsive. All serial I/O happens off the render thread.

Layout

The GUI is divided into five sections displayed top-to-bottom. A fixed status bar sits above the scrollable content and is always visible, and a safety toggle sits at the top of the scrollable area.

Safety Toggle

  • Safety: ON / OFF button at the top of the window — green when interlocks are active, red when bypassed
  • Clicking it flips and persists the state (shared with the CLI, survives restarts)
  • While OFF, a red "⚠ Safety OFF — start interlocks are bypassed" banner is shown
  • See Safety Interlocks for what the interlocks gate

Status Bar

  • Always-visible strip at the very top showing a one-line system summary
  • Five indicator chips: Compressor, ADR Ramp, GL7, Recording, Heatswitch
  • Chip is green (active) or dark/muted (inactive)
  • GL7 is considered active if any heater/switch output is non-zero or a cooldown subprocess is alive
  • Heatswitch is considered active when open

Thermometry

  • Live temperature readings from all LS350 inputs (4K Stage, ADR, 4-Switch, 3-Head, 4-Head, 3-Pump, 4-Pump) and LS370 input 1 (Device Stage)
  • Inputs using custom calibration (3-head, 4-head, pump and switch diodes) show calibrated temperatures in Kelvin
  • Record Temperatures button starts continuous CSV logging to temps/YYYY-MM-DD_temperature_log.csv at 30-second intervals; button turns red and reads "Stop Recording Temperatures" while active
  • Recording state persists across restarts (temps/.recording_active lock file). The GUI detects an interrupted recording on startup and resumes it automatically

Compressor

  • Start/Stop buttons with live status display (running state, error flags)
  • Compressor intent is persisted to disk (state/.compressor_intent). If FROST is restarted, the GUI restores the last known compressor state

ADR Cooldown

  • Live readouts: output current (A), voltage (V), magnetic field (T)
  • Input fields for ramp rate (A/s), target current (A), and soak time (minutes)
  • Start ADR Cooldown launches the full automated ramp sequence; button turns red and shows elapsed time while the ramp is running
  • Live log output displayed in the GUI during the ramp
  • Ramp state persisted to state/.adr_ramp_running to detect interrupted ramps on restart
  • Crash recovery: if the GUI starts and finds a stale ramp lock file (subprocess died), it reads the LS625 current and automatically ramps to 0 A if the magnet is still energized above 0.5 A

GL7 Sorption Cooler

  • Start GL7 Cooldown launches the full automated GL7 cooldown sequence (frost gl7 cooldown) as a subprocess; button turns red and reads "GL7 Cooldown In-Progress" until the process exits
  • CSV path field: paste the path to the active temperature recording, or click Current Temperature Recording to auto-fill from the running recording
  • Live output percentage display for all four outputs (4-Pump Heater, 3-Pump Heater, 4-Switch Heater, 3-Switch Heater)
  • DragValue + Set buttons to manually override any output percentage

CLI Reference

frost <device> [OPTIONS] <command> [ARGS]

Run frost --help for the full list of devices and commands, or frost <device> --help for device-specific options.

Common Device Commands

Device Common commands Notes
frost adr ramp <rate> <current> [--soak-mins N], logging Full automated ADR ramp sequence
frost compressor status, start, stop, all Cryomech compressor control
frost lakeshore625 current, set-current <A>, set-rate <A/s>, quench-status, raw Magnet power supply
frost lakeshore370 kelvin <input>, resistance <input>, set-heater <pct>, raw Device stage thermometry
frost lakeshore350 all, read <input>, set-output <N> <pct>, query-all-outputs, raw Stage thermometry + GL7 outputs
frost heatswitch open, close, home, stop, estop Zaber stepper motor
frost record-temps snapshot, loop [--interval N] Temperature CSV logging
frost safety status, on, off See Safety Interlocks

All Lakeshore devices accept --port and --baud overrides. The compressor also accepts --addr, the heatswitch accepts --device-id.

frost adr ramp — ADR ramp sequence

frost adr ramp 0.005 9.44 --soak-mins 60

The ramp command runs the full automated sequence: start background logger → set ramp rate and target current → wait for current to reach target (within 0.04 A) → soak at constant current → open heat switch + 3 min wait → ramp to 0 A → stop logger.

frost heatswitch — open/close behavior

open returns immediately — the motor executes asynchronously. close is blocking: it drives CCW until the home limit switch fires (position freezes at 0), polling until position stops changing or a 60-second timeout elapses.

frost lakeshore350 — input reference

Input Sensor Calibration
A 3-head thermometer CSV (Ω → K)
C 4-head thermometer CSV (Ω + 34.56Ω offset → K)
D2 4-switch diode CSV (V → K)
D3 4K stage diode LS350 curve 21 (K direct)
D4 3-pump diode CSV (V → K)
D5 4-pump diode CSV (V → K)

frost record-temps — temperature logging

frost record-temps loop              # Log at 30 s intervals (run in tmux)
frost record-temps loop --interval 60
frost record-temps snapshot          # Single reading to stdout

Logs are written to temps/YYYY-MM-DD_temperature_log.csv (17 fixed-width space-padded columns covering all LS350 inputs and LS370 input 1 device stage). Ctrl+C ends the recording. It is recommended to run this in a tmux pane if using CLI instead of the GUI.


GL7 Sorption Cooler Cooldown

The frost gl7 commands automate the cooldown of the Chase Research Cryogenics GL7 two-stage helium-3/helium-4 sorption cooler from 4K stage temperature (~3.6K) to ~310mK base temperature. The controller reads temperature data from the CSV written by frost record-temps loop and writes output percentages to the Lakeshore 350. Temperature recording must be running before and during the cooldown.

Output map

Output Connection Type Role
Output 1 4-pump heater Heater range 5 Heats the ⁴He pump
Output 2 3-pump heater Heater range 5 Heats the ³He pump
Analog Output 3 4-switch heater Analog Opens the ⁴He heat switch
Analog Output 4 3-switch heater Analog Opens the ³He heat switch

CLI commands

All subcommands require --csv pointing to the temperature log written by frost record-temps loop. Run frost gl7 --help for the full list of options and per-phase flags.

# Run the full automated sequence (Phase 0 → 5)
frost gl7 cooldown --csv temps/YYYY-MM-DD_temperature_log.csv

# Or run individual phases (each accepts --outN flags for manual handoff)
frost gl7 check      --csv temps/...   # Phase 0: verify outputs at 0%
frost gl7 ramp-pumps --csv temps/...   # Phase 1: ramp up both pumps
frost gl7 stabilize  --csv temps/...   # Phase 2: hold pumps, wait for heads to cool
frost gl7 cycle-4he  --csv temps/...   # Phase 3: cycle ⁴He module
frost gl7 cycle-3he  --csv temps/...   # Phase 4: cycle ³He module
frost gl7 running    --csv temps/...   # Phase 5: monitor at base temperature

Phase sequence

Phase 0: Output check  (~instant)
Phase 1: Ramp up both pumps  (~25 min)
Phase 2: Stabilize pumps, heads cool  (~60–90 min)
Phase 3: Cycle ⁴He module  (~60–75 min)
Phase 4: Cycle ³He module  (~25–35 min)
Phase 5: Running at base temperature  (~36 hours)

Total time to base temperature: ~3 hours. The full sequence can be run unattended with frost gl7 cooldown, or each phase can be run individually if manual intervention is needed between steps.

Phase details

Phase 0 — Output check: Verifies all LS350 outputs are at 0% before starting. Thermal preconditions are enforced by the safety interlocks (compressor running + 4K stage < 4.2K) at the point the cooldown is launched.

Phase 1 — Ramp up: Executes a fixed time-based ramp schedule (30%→50%→80% for Output 1, 30%→50%→60% for Output 2 over 90 seconds), then holds at 80%/60% and polls every 30 s. Once the 4-pump reaches 45K, Output 1 is stepped down by 8% per poll until it reaches 25%. Once the 3-pump reaches 42K, Output 2 is stepped down by 8% per poll until it reaches 18%. The two pumps step down independently. Phase 1 exits when both outputs have reached their floors (25% / 18%).

The 3-pump step-down starts at 42K, which is 3K below the target band of 45–55K. This is intentional overshoot prevention: at 60% output the pump is rising at several K/min, so waiting until 45K to begin stepping down would result in the pump coasting well past 55K before the reduced output takes effect. Starting at 42K gives the step-down ~3K of runway to arrest the rise before the pump enters the target range. The same principle applies to the 4-pump, whose step-down begins at 45K — 15K below its 60K upper limit.

Phase 2 — Stabilize: Holds 4-pump at 50–60K and 3-pump at 45–55K using a rate-limited feedback loop (adjustments at most once every 3 minutes per output). Uses rolling averages and dT/dt to avoid reacting to noise. Exits when the 4-head plateaus below 5.45K and both pumps have been in range for 10+ continuous minutes. Timeout exit available after 120 minutes if both heads are below 6.0K.

Phase 3 — Cycle ⁴He: Turns off 4-pump (Output 1 → 0%), opens 4-switch (Output 3 → 40%). Two concurrent control loops run every 30 seconds for the rest of the run:

  • 4-switch regulation (phases 3–5): keeps Switch_Temp_K (4-switch temperature) in 20–22K by adjusting Output 3 in the range 20–45% (−5% if above 22K, +5% if below 20K). If the switch has not reached 20K after 15 minutes a warning is logged; the feedback will keep increasing Output 3 toward 45% until it opens.
  • 3-pump management: Output 2 is boosted without a rate-limit if the 3-pump cools — the switch opening creates a large thermal disturbance that can rapidly cool the 3-pump. Includes predictive lookahead: if the rolling-average temperature extrapolated forward would drop below 45K within 1 minute (+8% boost) or 2 minutes (+5% boost), Output 2 is increased pre-emptively while the pump is still in range. Once the pump has already dropped below 45K the response is +8% (falling fast, < −0.3 K/min) or +3% (falling, < −0.1 K/min); below 40K the response is +10% (emergency). Output 2 can reach 100% during this phase.

Exits when both heads fall below 2.0K.

Phase 4 — Cycle ³He: Turns off 3-pump (Output 2 → 0%), opens 3-switch (Output 4 → 40%). Output 3 (4-switch) continues to be regulated in the 20–22K range as in Phase 3. No other adjustments are made. Exits when 3-head sustains below 350mK for 5 minutes.

Phase 5 — Running: Output 3 (4-switch) continues to be regulated in the 20–22K range. All other outputs are held at their Phase 4 exit values. Monitors every 5 minutes. Exits and alerts when ⁴He is exhausted (4-head > 3K and rising at > 0.01 K/min). Typical hold time: ~36 hours.

Safety limits

These override all phase logic at every iteration:

Condition Action
Any pump > 65K Reduce that pump's output by 20% immediately
4K stage > 12K Reduce all heater outputs by 10%
Phase 2 running > 180 minutes Log error, halt
Any output > 100% Clamp to 100%
Any output < 0% Clamp to 0%

Typical workflow

GUI: Click Record Temperatures to start logging, then in the GL7 section click Current Temperature Recording to auto-fill the CSV path and Start GL7 Cooldown.

CLI: Start temperature recording in one tmux pane (frost record-temps loop), then run frost gl7 cooldown --csv temps/... in another. Individual phases can be run separately if manual intervention is needed between steps.


Safety Interlocks

FROST enforces start-safety interlocks before any potentially dangerous operation is started. Two preconditions must both hold:

  1. The compressor is running.
  2. The 4K stage (LS350 input D3) is below 4.2 K (a reading ≥ 4.2 K blocks).

What is gated

Operation CLI GUI
ADR ramp frost adr ramp … Start ADR Cooldown button
GL7 full cooldown frost gl7 cooldown … Start GL7 Cooldown button
GL7 cold-start ramp (Phase 1) frost gl7 ramp-pumps …
Manual LS350 output ON frost lakeshore350 set-output <N> <pct> (pct > 0), outputs-set-range <N> <range> (range ≠ 0) GL7 output Set buttons (percentage > 0)

Not gated: read-only queries; turning an output off (0% / range 0); frost adr logging; and the mid-sequence GL7 phases (stabilize, cycle-4he, cycle-3he, running), which resume an already-started cooldown where the 4K stage is expected to be above 4.2 K.

Key behaviors

  • Start-only. The interlock is checked once, at the moment a process starts. It is never re-evaluated during a run — if the 4K stage rises above 4.2 K while an ADR ramp or GL7 cooldown is in progress, the process is not stopped.
  • Fail-safe. If a required reading cannot be obtained (serial error, sensor overrange, wrong port), the start is blocked, not allowed.
  • Persistent override. Safety defaults to ON and can be toggled OFF via frost safety off (CLI) or the Safety: ON / OFF button in the GUI. The state is stored in $HOME/.frost/safety_disabled, shared between CLI and GUI, and survives restarts. Use frost safety status to check the current state; frost safety on to re-enable.
  • Failure audit log. Every blocked start is announced on stderr and appended with a timestamp to logs/safety_log.txt. A blocked CLI command exits with an error naming the failed condition(s).

Data Logging

Both log types auto-increment filenames (_2.csv, _3.csv, …) when a file for the current date already exists.

Log Location Written by Interval
Ramp ramps/YYYY-MM-DD_ramp_log.csv frost adr ramp / frost adr logging 1 row/min
Temperature temps/YYYY-MM-DD_temperature_log.csv frost record-temps loop / GUI recording 30 s (configurable)
Safety logs/safety_log.txt Safety interlocks On blocked starts only

Temperature CSV columns (17 total, fixed-width space-padded): Timestamp, Date, Time, 4K_Stage_Temp, Switch_Volt, Switch_Temp, 3Head_Res, 3Head_Temp, 4Head_Res_Raw, 4Head_Res_Adj, 4Head_Temp, 3Pump_Volt, 3Pump_Temp, 4Pump_Volt, 4Pump_Temp, Device_Stage_Res, Device_Stage_Temp

Switch_Temp (Switch_Temp_K) is the 4-switch temperature (LS350 D2)


Dependencies

cryomech_api

compressor.rs uses the cryomech_api crate (SMDP V2 protocol) as a path dependency. It must be present at ../cryomech_api/ relative to the FROST directory. The cryomech_api crate itself requires smdp at ../smdp/.

# Clone both into the parent directory (sibling folders next to FROST/)
git clone <cryomech_api-repo-url> ../cryomech_api
git clone <smdp-repo-url> ../smdp

Expected directory layout:

parent/
├── FROST/
├── cryomech_api/
└── smdp/

If you do not have the compressor hardware and want to build without it, remove the cryomech_api dependency from Cargo.toml and the mod compressor; declaration from src/lib.rs and src/cli.rs.


Architecture Notes

Serial communication: All devices communicate over serial. Each command opens the port, executes, and closes immediately — there is no persistent connection. SCPI devices (LS625, LS370, LS350) use 7-bit odd parity framing; Zaber uses 6-byte binary frames; the compressor uses SMDP V2 binary protocol via cryomech_api.

Worker thread: The GUI spawns a background thread for all serial I/O so the render loop is never blocked. Device state is shared via Arc<Mutex<DeviceSnapshot>> and polled every 30 seconds.

Persistent state: The state/ directory holds lock files that survive process restarts. The GUI reads these on startup to restore compressor intent and detect interrupted ADR ramps.

Calibration files: The LS350 loads calibration CSVs (Ω→K or V→K interpolation tables) lazily on first use for the 3-head, 4-head, and pump diode sensors.


License

Apache 2.0

About

Fridge Remote Operations, Status, and Thermometry (FROST)

Resources

Stars

Watchers

Forks

Releases

Packages

Contributors

Languages